DIGITAL RADIOTHERAPY COBALT UNIT TERABALT
DIGITAL RADIOTHERAPY COBALT UNIT TERABALT
Digital Radiotherapy Cobalt unit TERABALT, the most modern product of UJP PRAHA, guarantees a high quality irradiation of oncological patients.
It has become the top product among cobalt irradiators thanks to the application of mechanical advantages of the previous generation of irradiators in combination with the development of the most modern digital technologies of control systems.
Its design benefits from full digitalization of all major elements – measurement, control, regulation and communication with operators and other external systems.
All movements are controlled by a digital sensor and operated by a separate module with its own processor.
All local control units are bidirectionally linked with the main therapy control panel that constantly monitors safety devices of the system and arranges the automatic setting of required parameters.
Operator has an easy access to control by means of a graphical user interface or a remote control pendant.
Both means of control make full use of automatic positioning to pre-set values transferred from the verification system or entered manually.
Medical Purposes
- Reliable and effective cobalt radiotherapy
- Outstanding spatial accuracy of the intended dose delivery
- Implementation of the most modern radiotherapeutic trends
- Wide spectrum, long-term stability and reliability of all functions
- Substantial increase in the number of treated patients
DIGITAL RADIOTHERAPY COBALT UNIT TERABALT
Control and Service
- Comfortable control
- Simple maintenanc
- On-line diagnostics of the unit’s operating conditions
- Easy calibration and dosimetry
Running Cost
- Minimal operating personnel requirements
- Negligible electricity consumption
- Minimal demands for the stability of electricity supply network
- Zero requirements for other operating media such as water or air
Low service costs (your trained personnel will manage most tasks)
Acquisition Cost
- Low acquisition value
- Minimal requirements for building adjustments of the workplace
- Simple and fast installation
- Easy upgrade option from SCS model to ACS model
The planning system is able to work both with cobalt irradiators and linear accelerators with a photon or an electron beam.
It is possible to import outputs of all diagnostic devices (supporting the output format of DICOM standard) as an input data.
A planning process consists of the following steps: selecting the patient and outlining the treated organs in various tomographic slices, setting suggested irradiation fields and calculating the absorbed dose in required amounts. The calculation and projection is always three dimensional.
The planning system is connected to the irradiator by data transportin DICOM RT format into the verification system.
You can also attach your own planning system (supporting DICOM RTformat) to TERAGIS verification system.
Your initial investment put into the purchase will thus be fully utilised.
An accurate dosimetry plays a key role in achieving the main target of radiotherapy – providing a maximal absorbed dose in the requested volume.
The irradiator is equipped with a separate irradiation mode for dosimetric purposes that allows to set all required parameters directly from the radiotherapy control panel without having to enter the irradiation room.
Additionally it is possible to arrange an automatic sequence of irradiation procedures for dosimetric purposes and thus to substantially shorten dosimetric measurements necessary for clinical operation
Extremely rigid construction of the device guarantees a fluent start, movement and positioning of all components.
Assembled stand with its gantry is firmly connected to the bottom frame which is concreted into the floor of the irradiation room.
The bottom frame does not require any additional anchoring.
The device is designed to be assembled in common conditions of standard irradiation rooms.
A transport route 120 cm wide is sufficient for the transportation of the device in dismantled state.
Mechanical adjustment of the system is necessary only after the assembly of its head
An updated collimation system irradiation beam is based on a brand new design.
DRCU TERABALT is equipped with a rectangular collimator working either in symmetric (SCS model) or asymmetric (ACS model) modes.
Both models are mechanically and electronically identical, they differ only in their controlling software thus allowing user to seamlessly switch from an economical version(symmetriccollimator)to a fully asymmetric collimation system without costly replacement of mechanical parts of the device.
The ranges of dimensions of the irradiation field are from 4 cm x 4 cm to 36 cm x 36 cm for type 80 and from 5 cm x 5 cm to 40 cm x 40 cm for type 100.
A DRCU TERABALT is equipped with a wide range of safety functions minimizing any potential danger to patients and operating personnel.
The basic safety component is an automatic securing of the return of the source to the shielded position based on a unique spring mechanism.
Secure power supply of safe 24V is another important safety component.
Moreover, the power supply is continually backed up, so an immediate supply failure does not lead to unexpected situations and the irradiation of the patient can always be completed without any problems.
All movements are locally controlled by independent electronic modules – regulators.
Accurate digital sensors ensure the highest possible precision of positioning to specified position.
This function together with the verification system minimizes potential mistakes of the operator.
The positioning of patients is ten times more accurate than the requirements specified by international standards.
Local independent control of each movement enables parallel operation of all mechanical parts noticeably contributing to faster setting of parameters.
DRCU TERABALT collimator is equipped with HW anti-collision bars that automatically block any movement when pressed.
For higher safety demands during irradiation it is possible to also supply software-based anti-collision system that continuously controls the mutual positions of all elements and starts an acoustic alarm in case any of the elements is breaching the safe distance.
Any movements following the alarm are allowed only from the remote pendant connected to the table in the irradiation room since direct contact between operator and the patient is required in such cases
TERAGIS verification system is afullyintegrated part of DRCU TERABALT.
All valid parameters of irradiation plan are transferred into it.
The verification system contains a calendar scheduling an optimum operation of the device.
The calendar works with patient’s data and irradiation fractions that the specific patient is supposed to obtain. I
f parameters are set incorrectly, TERAGIS prevents carrying out the irradiation procedure.
The communication between DRCU TERABALT and TERAGIS verification system is bidirectional, the verification system is informed about the condition of the device and the amount of completed irradiation time.
The verification system can be operated by multiple users.
For example, one operator can enter information about new patients and their schedule of irradiation procedures while the other operator controls irradiation process and another operator monitors the utilization of the device and prints out summary reports.
The exception is the control of the irradiation sequence as it can be carried out only from a workstation specified for that purpose.
DRCU TERABALT is equipped with wide spectrum of accessories.
Each accessory is marked with an individual code. After installation, the operating system identifies the accessories according to their codes.
If the identification code does not correspond to the code required, it is not possible to irradiate the patient.